Calibration of hplc in pharma

(1 )Measurement of Flow ml/min
this is measurement by flow meter or Cylinder
this is 0.3 ml, 0.5 ml ,1.0ml, 2.0ml ,3.0 ml
(2) Injector / auto sampler linearity test
Correlation coefficient NLT 0.990
(3) Column Compartment Temperature Test:
15 , 30 45 60
(4) Sample Compartment temperature test
5, 10, 15, 25
(5)Drift and Noise test
Drift  +- 2000 AU/hrs
Noise +- 100 AU
(6) Gradient test
Height (A/B/C) / Height of Full scale peak X 100      Limit 9.0 % – 11.0 %
(7) Pristion
NMY 2.0 %
(8) Detector
NLT 0.9990
205 245 273 333